Operation
H25-IR
30
Instruction 3015-5286
4.2.3 Range Selection
The H25-IR has three sensitivity ranges: SM (small), MED (medium), and LG (large). The range used when
searching for a leak depends on the size of the leak and the type of gas being monitored.
Press the
RANGE
button until the desired range is displayed on either the instrument’s Data Display screen, or
on the LCD of the Smart probe.
Instrument Display
Smart Probe Display
NOTE:
The range setting has no affect on the accuracy of the gas reading as displayed on the instrument when
making a leak-rate measurement. It does, however, change the sensitivity of the bar graph at the bottom of the
Data Display screen, the audible tone frequency as heard from the speaker, and the blink rate of the red LED
on the hand-held probe.
The “SM” range is the highest sensitivity level, and is most likely used for fluorine based gases like R134a and
most fluorine based refrigerant blends.
The “MED” range is used for chlorine and chlorine based gases like R12, R22, R123, etc. This position is also
useful for locating large R134a or other fluorine based refrigerant leaks of 1
2 Oz/yr.
The “LG” range is used to locate very large leaks.
TIP:
Blowing out the test area with shop air may enable the operator to locate a large leak more quickly.
4.2.4 Setup Selection
There are four user-defined Setups that allow the operator to quickly switch between frequently used
combinations of Gas Type, Range, Feedback Mode, Setpoint, and the Auto Zero Off condition (refer to Section
2.5 Configuring the SETUPS Function
).
When first turned ON, the instrument returns to the last Setup selected. To select a different Setup, first press the
SETUP
button to display the Setup Selection menu. Next, use the
Up
and
Down
buttons to highlight the desired
Setup, and then press
ENT
to make the selection and be returned to the Data Display screen. The name of the
newly selected Setup will appear on the Data Display screen.
NOTE: The names listed in the Setup Selection menu are user defined, and may vary from what is
shown above. The number of each setup, however, is what appears on the display of the Smart probe.
Instrument Display
Smart Probe Display
SETUP NO. 1
SETUP NO. 2
SETUP NO. 3
SETUP NO. 4
Setup Name
and Number
0.00 SM
SRCH (1)
Oz/yr
R134A
SETUP NO
SM SRCH
Range: SM, MED, LG
0.00 SM
SRCH (1)
Oz/yr
R134A
SETUP NO
SM SRCH
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...