Calibration
Instruction 3015-5286
27
3.5 Restoring Factory Calibration
NOTE: Factory calibration cannot be restored if the instrument was calibrated using an external leak source.
Restore an instrument to its factory calibration value as follows:
Display the Calibration Data menu by first pressing
both
the
ENT
and
ESC
buttons on either the instrument’s
front panel or Smart probe at the same time to display the Main menu. Next, use the
Up
and
Down
buttons to
highlight “CAL DATA”, and then press
ENT
to display the Calibration Data menu.
Main Menu:
SYSTEM
LOG
SETUPS
CAL
DATA
DIAGNOSTICS
From the Calibration Data menu, use the
Up
and
Down
buttons to highlight “RESTORE FACTORY CAL”, and
then press
ENT
to display the RESTORE FACTORY CAL screen.
Press the
ENT
button to confirm restoring the instrument to its factory calibration values.
3.6 Storing a New Factory Calibration
For all models fitted with an internal leak source, "factory" calibration data is stored in memory and used to
validate all subsequent internal calibrations performed by the user. An internal calibration failure will occur when
the calibration data collected during an internal calibration cycle, performed by the user, deviates from the
"factory" calibration data. This can occur when the internal particulate filter needs to be replaced or if a probe is
changed. When the internal filter or probe is replaced, or the optical sensor, pump or pneumatics are serviced,
a new "factory" calibration must be stored. This is accomplished by first performing an internal calibration, which
may fail due to the changes to the instrument, then pressing the ENT and CAL keys together.
Press ENT to confirm. This will take the data collected from the calibration just completed and store it as the
"factory" or reference calibration for validation of all future internal calibrations.
Calibration Data Menu:
CAL TYPE
EXT
RATE
EXT
GAS
TYPE
CAL
FACTOR
RESTORE FACTORY CAL
CAL INTERVAL
ENFORCE CAL INTERVAL
LAST CAL DATE
RESTORE FACTORY CAL
HIT ENTER TO CONFIRM
PRESS ENTER TO
STORE LAST CAL
AS REFERENCE
g/yr
FAULT
SETUP NO.1
SM SRCH
LAST CAL HAS
BEEN STORED
AS REFERENCE
g/yr
FAULT
SETUP NO.1
SM SRCH
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...