H25-IR
Preparing for Operation
Instruction 3015-5286
15
2.4.8 Serial Output
Leak rate data can be output as ASCII text in two formats.
After selecting the SERIAL OUTPUT function, the user can choose to have the text output to a file in two
formats.
Refer to section 4.6.
This function allows the string to be written to a file at the rate of 10 times per
second.
2.4.9 Probe Button
The PROBE BUTTON function selects what action is performed by the pushbutton on the Standard Probe with
LED and Pushbutton.
DISABLED
– No action performed.
MODE KEY
– Toggles the instrument between its Search and Measure modes.
ENTER KEY
– Performs the same function as the
ENT / ZERO
key on the instrument’s front panel.
Use the
Up
and
Down
buttons to highlight the desired setting. Press
ENT
to save the selection.
2.4.10 Reset Defaults
The RESET DEFAULTS function returns the instrument to its factory default settings per Appendix A.
This operation requires a three key press. Simultaneously press the ENTER+VOL buttons to reset
defaults.
2.5 Configuring the SETUPS Function
There are four user-defined Setups that allow the operator to quickly switch between frequently used
combinations of Gas Type, Range, Feedback Mode, Setpoint, and the Auto Zero Off condition. Program the
four Setup configurations as follows.
First enter the Setups menu by pressing
both
the
ENT
and
ESC
buttons on either the instrument’s front panel or
Smart probe at the same time to display the Main menu. Next, use the
Up
and
Down
buttons to highlight
“SETUPS”, and then press
ENT
to display the Setups menu.
Main Menu:
SYSTEM
LOG
SETUPS
CAL
DATA
DIAGNOSTICS
NOTE: The names listed in the Setups menu are user defined, and may vary from what is shown above.
Setups Menu:
SETUP NO.1
SETUP
NO.2
SETUP
NO.3
SETUP
NO.4
Configuration Menu:
SETUP NAME
RANGE
GAS TYPE
FEEDBACK MODE
SETPOINT
SEARCH AUTO ZERO
CONFIRM RESET TO
DEFAULTS BY PRESSING
THE UP ARROW
PROBE BUTTON
DISABLED
MODE KEY
ENTER KEY
SERIAL OUTPUT TYPE
NONE
TEXT PACKET
TEXT ONLY
<CLOCK
PCHK MODE
SERIAL OUTPUT
<PROBE BUTTON
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...