Preparing for Operation
H25-IR
16
Instruction 3015-5286
From the Setups menu, use the
Up
and
Down
buttons to highlight the desired Setup. Press
ENT
to select that
Setup and display its Configuration menu.
From the Configuration menu, use the
Up
and
Down
buttons to highlight the desired function. Press
ENT
to
select that function and display its setup screen. Perform the appropriate setup procedure as described in
Sections 2.5.1 thru 2.5.6.
2.5.1 Setup Name
The SETUP NAME function allows the operator to give each Setup a unique name that will appear on the Data
Display screen. Use the
Left
and
Right
buttons to move the blinking cursor across the screen to highlight the
character to be modified, and then use the
Up
and
Down
buttons to modify that character. When finished
entering the desired characters, press
ENT
to save the new name.
2.5.2 Range
The RANGE function sets the sensitivity of the gas-level graphical display (bar graph located at the bottom of
the display), audible tone frequency, and the hand-held probe’s LED blink rate in relation to the detected gas
level. Use the
Up
and
Down
buttons to highlight the desired sensitivity. Press
ENT
to save the selection.
NOTE:
The range setting has no affect on the accuracy of the gas reading as displayed on the instrument when
making a leak-rate measurement.
2.5.3 Gas Type
The GAS TYPE function sets the type of gas being monitored. Use the
Up
and
Down
buttons to highlight the
desired gas. Press
ENT
to save the selection.
NOTE:
The gases displayed depend on which optical bench is installed. Refer to Section 6.
2.5.4 Feedback Mode
The FEEDBACK MODE function provides the operator both visual and audible feedback in regard to the
measured leak rate. This function determines how the gas-level graphical display (bar graph located at the
bottom of the display) and audible tone operate as follows.
CONTINUOUS
– The bar graph and audible tone continuously increase and decrease their values in step with
the leak rate. For example, when the leak rate increases, the bar graph moves upscale and the audible tone
increases in frequency. Their maximum values at various leak rates are determined by the Range setting.
SETUP NAME #1
S
ETUP NO. 1
RANGE LG
SETUP NO. 1
MED
SM
GAS TYPE
<
R134A
SETUP NO. 1
R401A
R402A
<
R402B
FEEDBACK MODE
SETUP NO. 1
CONTINUOUS
SETPOINT
H25C
SETPOINT
Summary of Contents for H25-IR
Page 6: ...Contents H25 IR iv Instruction 3015 5286 Notes ...
Page 26: ...Preparing for Operation H25 IR 20 Instruction 3015 5286 Notes ...
Page 34: ...Calibration H25 IR 28 Instruction 3015 5286 Notes ...
Page 43: ...H25 IR Operation Instruction 3015 5286 37 Notes ...
Page 68: ...Specifications H25 IR 62 Instruction 3015 5286 Notes ...