Instruction 0024-9395
43
Fyrite Pro
3. Set the simulator to 32.0 °F (0 °C); then use the
▲
and
▼
buttons to
set the dis played value to match the simulator’s value.
4. Set the simulator to 212.0 °F (100 °C); then use the
▲
and
▼
but-
tons to adjust the dis played value to match simulator’s value.
5. Repeat Steps 3 and 4 as necessary until the unit is reading cor-
rect ly at both temperatures.
6. Press
ENTER
to store these calibration values and move to the next
cal i bra tion screen, or hold down
ENTER
for 2 seconds to store these
calibration values and enter the Run Mode, or press the
I/O
button
to exit the Cal i bra tion Mode and turn OFF the an a lyz er with out
saving the changes.
7. Remove thermocouple simulator from analyzer.
5.4 Stack Thermocouple Calibration
(Model 110/120/125)
Material Required:
Thermocouple Simulator
Procedure:
1. Attach the thermocouple simulator to the Stack Ther mo cou ple con-
nector on the analyzer (T-STACK, see Figure 3-2).
2. Enter the Calibration Mode as described in Section 5.1. Then re-
peatedly press the
ENTER
button until “STACK” appears at the top
of the display.
3. Set the simulator to 32 °F (0 °C); then use the
▲
and
▼
buttons to
set the dis played value to match the simulator’s value.
4. Set the simulator to 572 °F (300 °C); then use the
▲
and
▼
buttons
to adjust the dis played value to match simulator’s value.
5. Repeat Steps 3 and 4 as necessary until the unit is reading cor-
rect ly at both temperatures.