Instruction 0024-9395
28
Fyrite Pro
4.9 Clearing Saved Test Data
(Model 105/110/120/125)
Clear all saved test data as follows:
Note:
If no test data has been saved, the option to clear
memory will not be available.
1. With the analyzer in its Run Mode, press the
HOLD
button
twice
to
display the Print Screen.
2. Press the
▲
or
▼
button to scroll to the Clear Screen, and then press
ENTER
to display the Clear All Screen.
3. Press
ENTER
again to clear memory and return to the Hold Mode, or
press
HOLD
to return to the Hold Mode without clearing memory.
4.10 Turning OFF the Analyzer & CO Purge
Press the
I/O
button to turn OFF the analyzer. The unit will count down
from 5 before turning OFF, thus allowing time for the operator to abort the
turn OFF process by press ing the
ENTER
button.
If a high CO level is detected at turn OFF, the unit will remain ON with
its pump running and display “PUrG CO”. The countdown from 5 will not
begin until the detected CO level drops below 50 ppm. Although not recom-
mended, the purging process can be bypassed by pressing the
I/O
button a
second time.