T E C H N I C A L D ATA
Type of unit
Volumetric infusion pump
Classification (acc. to IEC/EN 60601-1)
defibrillator-proof; CF equipment
Protective Class II; Protective Class I in
combination with SpaceStation
Class (acc. to Directive 93/42 EEC)
IIb
Moisture protection
IP 22 (drip protected for horizontal usage)
External power supply:
•
Rated voltage
Via B. Braun SpaceStation or optional
mains adaptor (rated voltage 100 … 240 V
AC
~
, 50/60 Hz) for stand alone operation
•
External low voltage
11 … 16 V DC
via Connection Lead SP
12 V or via SpaceStation
Staff call
Max. 24 V / 0,5 A / 24 VA (VDE 0834)
EMC
IEC/EN 60601-1-2 / 60601-2-24
Time of operation
100 % (continuous operation)
Operating conditions:
•
Relative humidity
30 % … 90 % (without condensation)
•
Temperature
+10 … +40 °C
•
Atmospheric pressure
500 … 1060 mbar
Storage conditions:
•
Relative humidity
30 % … 90 % (without condensation)
•
Temperature
-20 … +55 °C
•
Atmospheric pressure
500 … 1060 mbar
Type of battery pack (rechargeable)
NiMH
Operating time of rechargeable battery
Approx. 4 hours at 100 ml/h
Recharging time
Approx. 6 hours
Weight
Approx. 1.4 kg
Dimensions (W x H x D)
214 x 68 x 124 mm
Volume preselection
0.1 - 99.99 ml in increments of 0.01 ml
100.0 – 999.0 ml in increments 0.1 ml
1000 – 9999 ml in increments 1 ml
Time preselection
00:01 – 99:59 h
Accuracy of set delivery rate
± 5 % according to IEC/EN 60601-2-24
Alarm in the case of incorrect dose
For incorrect dosages of 1,4 ml due to
malfunctions of the device the pump will
automatically shut off
Technical inspection (safety check)
Every 2 years
TECHNICAL DATA