total dose and the remaining time. In the Main Menu total dose, time and VTBI
can be changed, also during RUN-Mode. Other parameters (dose rate, basal rate,
concentration, patient weight and patient height) cannot be changed.
Note:
The KVO function and Bolus function are disabled during Dose Over Time.
Note:
The feature Dose Over Time always requires the usage of dosing units (i.e.,
mg or mg/kg patient weight).
Before using Dose Over Time contact your local B. Braun representative!
Starting Dose Over Time via Drug Library:
Note:
Dose Over Time settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable and use the drug library according to the Instructions for
Use.
•
Select a drug by using
t
and press
l
.
The pump now offers the possible therapy profiles. Select “Dose over Time” with
t
and
press
l
.
The editor for Total Dose is shown if a drug with therapy Dose over Time is sel-
ected from drug library and no default value for Total Dose was entered in libr-
ary. The editor is also shown if the Total Dose is edited in the Main menu.
The editor for Time is shown if a drug with therapy Dose over Time is selected
from drug library and no default value for Time was entered in library. The editor
is also shown if the Time is edited in Main Menu.
The VTBI is calculated automatically and the following screen is displayed:
Enter the total dose, if necessary,
and confirm with
k
.
Enter the time, if necessary,
and confirm with
k
.
Check calculated rate by using
t
for plausibility
Start Dose Over Time by
pressing
s
f
.
Chapter 3
SPECIAL FUNCTIONS
42