51
Check the line for damage and make sure that the
line is correctly inserted.
"Flow"
Drop chamber is completely filled or leak in the
system.
Examine the line for damage and check the drop
chamber.
“Data were reset”
Therapy and pump settings could not be restored.
Enter therapy again.
"Therapy data were reset"
Therapy data could not be restored.
Enter therapy again.
“Data Lock"
An attempt was made to stop or switch the pump off
without entering the code.
Enter the correct code in order to continue the
therapy respectively turning the pump off.
The red LED doesn’t extinguish until the administration is started again respectively the
pump is turned off.
Caution:
If a wrench is displayed and/or a yellow, red and blue LED blink then the
pump is in the service mode and is not permitted to be used on a patient. The pump
must then be checked by a service technician.
5.3 Reminder Alarms
Reminder alarms only occur in two cases:
1. A line is inserted, the pump does not deliver, no value is edited and the device is not
operated for two minutes.
An acoustic tone sounds, the yellow LED blinks and a staff call is activated.
a) The display states “Reminder alarm!”
b) The display states “Config. not finished!”
Confirm alarm with
K
and continue to set therapy/Start Up configuration.
2. A value edition was started but not finished and confirmed. This is also possible
with a missing disposable.
An acoustic tone sounds, the display states “Value not accepted“, the yellow LED
blinks and a staff call is activated.
Confirm alarm with
K
and continue to set therapy.
ALARMS
Chapter 5