24
Chapter 3
SPECIAL FUNCTIONS
The therapy can be started now with
s
f
in case all values are defined.
Depending on the pre-defined settings the therapy is started with an initial
bolus and a basal rate or not.
Before leaving the patient the pump should be put into DataLock level 3 with
q
in Menu “Options”. This is necessary especially in case non-authorised access
to the settings can be anticipated.
The code is entered with
q
and confirmed with
k
.
In this state the patient is allowed to demand boli. Depending on the status of
the therapy these are either administered or denied. Changing the disposable is
also possible by using the code for level 1 or level 2. Altering the settings for
PCA or other therapies however is only possible with the code for level 3.
The status of the therapy can be checked in the menu „Status“ .
Enter the „Main Menu“ with
c
and select the “Status“ with
q
.
An acoustic confirmation of demanded boli can be activated and modulated by
t
in Data Lock 3.
Is a demand button connected, the therapy symbol looks like this:
In case there is no demand button connected the therapy symbol looks like
this:
.
The demand button is connected to the interface P2 at the rear side of the
pump.
The A/D-ratio indicates the
percentage of administered and
demanded boli thus giving an idea
about the effectivity of the therapy.
The pump display now may look
like this.
Select profile PCA by using and
confirm with
t
.The therapy
settings stored in the drug list are
displayed *.
After the selection the pump
offers additional drug related
Information which are confirmed
by
l
.
*Bolus volume is the volume of a single bolus the patient may demand. Max. Limit is the amount of
drug or volume a patient may demand within a certain time in total. Lockout is the time in between
two boli.