Chapter 3
SPECIAL FUNCTIONS
40
Starting Intermittent Mode via Drug Library:
Note:
Intermittent Mode settings have been configured in the Drug List
Manager before and have been uploaded into the pump.
•
Switch on pump with
o
and wait until self-check is finished.
•
Insert disposable and use the drug library according to Instructions for Use.
•
Select the desired drug with
t
and press
l
.
The pump now offers the possible therapy profiles.
•
Select “Intermittent Mode” with
t
and press
l
. The therapy settings for
“Intermittent Mode” are shown on the display.
•
For changing the parameters, press
l
to change and
k
to confirm.
Note:
Bolus rate is calculated by editable parameters. These parameters have to
be checked by the user before starting the infusion.
The pump can be started now by pressing
s
f
.
Starting Intermittent Mode via Special Function Menu:
•
Switch on pump with
s
f
and wait until self-check is finished.
•
Insert disposable.
•
Go to Special Functions Menu and select Intermittent Mode.
•
Press
l
to enter parameters and
k
to confirm.
After entering all desired parameters the pump can be started by pressing
s
f
.
In the upper part of the display the icon for “Intermittent Mode” appears.
In bolus phase the screen shows the following:
After the bolus phase the pump switches to rate phase and the screen shows
the following:
Note:
To cancel bolus infusion in the Intermittet Bolus therapy at any time it is
only possible with
s
f
.
Note:
The delivery of drugs can be stopped and started again in the Intermittent
Mode at any time by pressing
s
f
.
The pump now delivers the
predefined bolus.
The pump now delivers the
predefined rate.