Chapter 3
SPECIAL FUNCTIONS
The infusion is within the range of the maximum soft limit
=
The infusion is within the range of the minimum soft limit
=
Violation of the upper soft limit
=
Violation of the lower soft limit
=
No soft limit is defined
=
Only a drug name is available
=
(It is possible to select a drug name only from the drug library)
The limits of the drug library have to comply with the limits of the pump and
the disposable.
Note:
An adequate monitoring when infusing highly potent drugs is
recommended.
Note:
In case a drug from the drug library is selected and the pump is running
under dose rate calculation the initial values will be overwritten by the drug
library values if selected.
Remote Drug Library update from Upload Manager (Space Online Suite)
The file icon blinks every 2 s. An update is available.
The Drug Library Upload starts as soon as the pump is in Passive mode.
Note: You can cancel the upload by pressing
c
.
Please contact your local sales represantative in case you like to use Remote
Drug Library update.
3.4 Patient Controlled Analgesia (PCA)
For PCA a drug list with at least one drug activating the profile PCA is
necessary. By this the conditions for an effective and safe therapy are defined.
Switch on pump with
o
and wait until self-check is finished. Depending on the
settings the choice of a drug is offered direcly or the pump is in “Main Menu”.
Select “Special Functions” with
t
from “Main Menu” and confirm with
l
.
Select drug list, category and desired drug by using
q
.