After loggings:
The logging will stop after the pre-defined number of readings has been reached.
By date/time
:
Logging is stopped on user-defined date and time
Interval
:
Defines the interval between readings: hour, minutes and seconds. When the logging mode is set to the
Instantaneous
the minimum interval is 1 second. For
maximum
and
minimum
values, the minimum interval is 10 seconds in
both models.
Misc
:
Shows the Logging Mode parameter that defines how measured value will be logged. Options are:
Instantaneous
:
The value logged will be exactly the value measured at each interval. Measurement is taken at the end of the
defined interval. The minimum interval between measurements is 1 second.
Average
The Logger takes 10 equidistant measurements within the predefined measurement interval (Period), build the
average of these ten values and store the result in the internal memory. The minimum interval between
measurements is 10 seconds
Minimum
:
The value logged will be the minimum value found within ten consecutives measurements taken within an
interval. The minimum interval between measurements is 10 second.
Maximum
:
The value logged will be the maximum value found within ten consecutives measurements taken within an
interval. The minimum interval between measurements is 10 second.
Alarms
:
Enables an alarm that is triggered according to user defined parameters. Values shown in the temperature unit
selected.
Clocks
:
Provides access to Logger and PDA clocks. When a new configuration is sent to the logger, clocks are
updated.
Offset
:
allows fine offset adjustment on the measured value. The
Reset
button clears changes made and the logged
values are the measured values again.
Click
Apply
to send this configuration to the Logger, returning to the
Monitoring
screen. Sending a new configuration implies the
exclusion
of all data collected present in the logger memory