4.3
Exchange of the Sensing Head
From factory side the sensing head has already been connected to the
electronics and the calibration code has been entered. Inside a certain model
group any exchange of sensing heads and electronics is possible. The sensing
heads and electronics of the
DM Fast
models (
LT151F
and
LT251F
) and DM
4M
cannot be exchanged.
4.3.1
Entering of the Calibration Code
Every head has a specific calibration code, which is printed on the head cable. For a
correct temperature measurement and functionality of the sensor this calibration code
must be stored into the electronic box. The calibration code consists of 3 blocks (1M, 2M,
3M = 5 blocks) with 4 characters each.
Example :
A6FG
–
22KB
–
0AS0
1.Block
2.Block
3.Block
For entering the code please press the Up and Down key (keep pressed) and then the
Mode key. The display shows HCODE and then the 4 signs of the first block. With Up and Down each sign can be changed; Mode switches to the
next sign or next block.
The entering of a new calibration code can also be made via the CompactConnect software (optional)..