Display
Mode [Sample]
Adjustment Range
142.3C
Object temperature (after signal processing) [142,3 °C]
fixed
127CH
Head temperature [127 °C]
fixed
25CB
Box temperature [25 °C]
fixed
142CA
Current object temperature [142 °C]
fixed
MV 5
Signal output channel 1 [0-5 V]
0-20
= 0–20 mA/
4-20
= 4–20 mA/
MV5
= 0–5 V/
MV10
= 0-10 V/
TCJ
=
thermocouple type J/
TCK
= thermocouple type K
E0.970
Emissivity [0,970]
0,100 ... 1,100
T1.000
Transmissivity [1,000]
0,100 ... 1,100
A0.2
Signal output Average [0,2 s]
A----
= inactiv/
0,1 … 999,9
s
P----
Signal output Peak hold [inactive]
P----
= inactiv/
0,1 … 999,9
s/
P oo oo oo oo
= infinite
V----
Signal output Valley hold [inactive]
V----
= inactiv/
0,1 … 999,9
s/
V oo oo oo oo
= infinite
u0.0
Lower limit temperature range [0 °C]
depending on model
/ inactive at TCJ- and TCK-output
n 500.0
Upper limit temperature range [500 °C]
depending on model
/ inactive at TCJ- and TCK-output
[ 0.00
Lower limit signal output [0 V]
according to the range of the selected output signal
] 5.00
Upper limit signal output [5 V]
according to the range of the selected output signal
U °C
Temperature unit [°C]
°C/ °F
| 30.0
Lower alarm limit [30 °C]
depending on model
|| 100.0
Upper alarm limit [100 °C] AL2
depending on model
XHEAD
Ambient temperature compensation [head temperature]
XHEAD =
head temperature/ -40,0 … 900,0 °C (for LT) as fixed value for
compensation/ returning to XHEAD (head temperature) by pressing
Up
and
Down
together
M 01
Multidrop adress [1] (only with RS485 interface) RS422 mode
01…32
RS422
(Press Down button on M01)
B 9.6
Baud rate in kBaud [9,6]
9,6/ 19,2/ 38,4/ 57,6/ 115,2
kBaud
S ON
Laser Sighting (3 VDC switch to connection pin 3V SW)SW)
ON/ OFF