OPEARTING MANUAL
Technische Änderungen vorbehalten
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
0141 0315-100 17.08.2015
Fon +49 771 83160 | Fax +49 771 831650 | [email protected] | bb-sensors.com
76 / 88
6.2
Error Messages
The display of the sensor can show the following error messages:
LT/ G5/ P3/ P7 models
Display
Meaning
OVER
Object temperature too high
UNDER
Object temperature too low
^^^CH
Head temperature too high
vvvCH
Head temperature too low
1M/ 2M/ 3M models:
1. Digit
Meaning
2. Digits
Meaning
0x
No error
x0
No error
1x
Head temperature probe short circuit to GND
x2
Object temperature too high
2x
Box temperature too low
x4
Head temperature too low
4x
Box temperature too high
x8
Head temperature too high
6x
Box temperature probe disconnected
xC
Head temperature probe disconnected
8x
Box temperature probe short circuit to GND