Azure Biosystems Cielo AIQ030 User Manual Download Page 22

Azure Cielo Real-Time PCR Systems User Manual

Page 22

 

   

2.   Time settings:

 

   

 

•   Tap on the time display button for each cycle to alter the time spent at that step. Time can 

be entered in either minutes or seconds. Enter the time using the key pad on the screen.

       

Example:

 To set the time to 2 minutes, select Min on the screen and enter the number 2 

using the key pad on the screen. Press DONE to save.

       

Example 2:

 To set time to 1 minutes 30 seconds, tap on Min on the screen and enter the 

number 1. Tap on Sec and enter the number 30 using the key pad on the screen. Press 

DONE to save.

 

   

 

•   Tap C   to clear the entered numerical digits.

 

   

3.   Go To module and Cycle Settings:

 

   

 

 The Go To module and the Cycle Settings are closely linked. Regardless of which step is 

highlighted, by tapping on the GO TO button, a two-step set of conditions will be added to the 

end of the run (but before the melt curve portion of the protocol). The first step is a 95ºC for 10 

seconds followed by a 60ºC for 10 seconds. A repeat cycling default of 30 cycles will also be 

visible in a small purple oval with an arrow indicating to which step of the cycle it will go after 

completion of the last step. All parameters are adjustable and can be changed. Additional 

steps may also be added if something other than a two-step protocol is required

 

   

 

•    Tapping the GO TO option will insert two additional steps at the end of the protocol. The 

cycles will be repeated for a default of 40 times. User can freely add or delete steps with 

the   and   buttons and adjust temperature and time for these new steps

 

   

 

•    Tap on the purple display button 

 at the end of the CYCLING portion of the RUN 

SETUP display. The number listed in this purple oval is how many times the conditions 

will be repeated. By tapping on the purple display, the number of repeats (cycles) and the 

GO TO location can be adjusted (GO TO which step). Be sure to tap the MODIFY button to 

save the changes.

 

   

 

•    The default setting is for two-step cycling at 95ºC and 60ºC. The default experiment is set 

with camera ON (to detect) at the end of the 60ºC step. Please note, if steps are deleted 

or added, the default camera setting is OFF. Please review the camera settings to ensure 

detection is ON and is happening at the appropriate step.

       

Example:

 To set 40 cycles consisting of step 2 and 3, tap on the purple button that 

displays number of cycles. A dialogue box will open, and the number of cycles can be 

adjusted as well as the step to GO TO.

Figure 24. Go To module on the 

RUN SETUP screen.

Summary of Contents for Cielo AIQ030

Page 1: ...Azure Cielo Real Time PCR Systems User Manual Part Numbers AIQ030 and AIQ060...

Page 2: ...d to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment...

Page 3: ...e Biosystems or the company s authorized agent Use of spare parts supplied by anyone other than Azure Biosystems Damage caused by accident or misuse Damage caused by disaster Corrosion caused by impro...

Page 4: ...um Informationen hinsichtlich der Entsorgung Ihres Ger tes zu erhalten ITA Questo simbolo indica che i rifiuti derivanti da apparecchiature elettriche ed elettroniche non devono essere smaltiti come...

Page 5: ...net Network 11 2 2 Connect Azure Cielo Real Time PCR system to PC via Azure Cielo Manager software 11 3 Device Software Overview 14 3 1 The Home Screen 14 3 2 Experiment Setup 14 3 2 1 Protocol Manage...

Page 6: ...10 50 L recommended 5 150 L 10 50 L recommended Excitation source LED LED Detection channels 3 6 Multiplexing Up to 3 targets Up to 6 targets Thermal element Peltier Peltier Max block ramp rate 6 C se...

Page 7: ...tic handling Azure Cielo Real Time PCR Systems cannot accommodate a fully skirted 96 well plate Azure Cielo Systems are compatible with only a non skirted or semi skirted 96 well plate Plate color can...

Page 8: ...ltra clear The Azure Cielo Systems are not compatible with domed caps for tubes or plates C Only a few samples While the Cielo Real time PCR system is designed for 96 well plates many times researcher...

Page 9: ...or call 925 307 7127 9am 4pm Pacific time Figure 2 Filled orange circles show locations of samples across the 96 well format There are a total of six samples in tubes Panel A shows an uneven distribu...

Page 10: ...be placed away from interfering electrical signals and magnetic fields If possible a dedicated electrical outlet should be used to eliminate electrical interference from other instrumentation in your...

Page 11: ...the SEARCH DEVICES button to locate available Azure Cielo Real Time PCR systems within the Wi Fi network Select the instrument to be used and then CONNECT The designated instrument and the PC should n...

Page 12: ...witch on the instrument using the power switch on the back Wait until the front power button light shows green Once green is visible on the front press and hold the front power button for eight second...

Page 13: ...e should immediately be removed from the USB port on the front Figure 9 Once the instrument has successfully updated the software it will instruct the user to remove the USB drive and select OK to reb...

Page 14: ...rotocol see Section 3 2 Data Manager Access run data see Section 3 3 Device Settings Access device settings see Section 3 4 Connection Bar Status and strength of internet connection Table 1 Buttons in...

Page 15: ...OTOCOL MANAGER screen under Experiment Setup Button Button Name Function Home Back to Home Screen Connection Bar Status and strength of internet connection Sort Sort in ascending or descending order A...

Page 16: ...can see the new user added to the PROTOCOL MANAGER screen Figure 13 Add New User process flow C Add New Protocol Tap symbol to add a new protocol Tap on Add Protocol to add a new user Enter the name f...

Page 17: ...delete all existing protocols and data stored under that user The system will prompt you to confirm the deletion Tap OK to delete or CANCEL to abort Figure 16 Delete exisiting user or protocol in PRO...

Page 18: ...THER button to customize reaction volume Lid Temperature Select lid temperature default setting is 105 C Off lid heating Turn off lid heating Light Source Select light source s Back Back to PROTOCOL M...

Page 19: ...empty if turned off see images below To deselect wells tap on the well to clear selection Wells are deselected in sets of two columns Press OK to save changes on to the plate and return to PROPERTIES...

Page 20: ...p and run protocol Email notification with results attached will be sent when the protocol run is successfully completed 3 2 5 Run Setup Figure 23 RUN SETUP screen Button Button Name Function Home Bac...

Page 21: ...s will delete the highlighted step Go To module Send the program to a cycling parameter in a previous part of the protocol Save This will save the Run settings Camera and or Gradient PCR Advance setti...

Page 22: ...f the last step All parameters are adjustable and can be changed Additional steps may also be added if something other than a two step protocol is required Tapping the GO TO option will insert two add...

Page 23: ...ra icon again and the icon will return to the inactive state light gray Alternatively within the step where detection needs to be active tap on the settings icon A dialog box will open with a toggle s...

Page 24: ...he MODIFY button to retain the new exposure setting or CANCEL to abort it Figure 25 Select channel to be adjusted The switch button turns BLUE when the camera is ON and GRAY when the camera is OFF The...

Page 25: ...olumn across the plate Users are not able to alter those assigned temperatures and a minimum difference of 1 degree is required for the temperature range The GRADIENT option can be accessed by opening...

Page 26: ...ssed over a short amount of time adjusting to a faster ramp rate could significantly shorten the overall run time If a particularly long or high GC content primer is in use a slower ramp rate might en...

Page 27: ...at 1 C this means for every increase of temperature of 1 C the plate will be scanned and data will be recorded Higher resolution means more scans are performed acquiring more data with each scan 0 1 C...

Page 28: ...re 31 The screen display after RUN button activated The run time elapsed and remaining can be monitored using the status box near the bottom of the run screen The green meter indicates how much of the...

Page 29: ...een displays the percentage of experiment completed and time remaining At the initiation of a run the instrument will provide a default experiment file name and a location to store the data The user c...

Page 30: ...tatus and strength of internet connection Sort Sort in ascending or descending order Delete Delete User Delete Protocol Delete Experiment Export Export Data Table 6 Buttons in the DATA MANAGER screen...

Page 31: ...ation curve window within the preview area The Amplification Curve data will be shown as seen in Figure 36 Color coded buttons to the left of the plot window correspond to all the channels available w...

Page 32: ...graphs to be visible or hidden To return to the DATA MANAGER screen close the viewing windows by clicking on the CLOSE option at the very top right of the plot window B Delete User From the listed us...

Page 33: ...xported to a USB drive using one of the two ports on the instrument Data can be exported via email To email files please verify the following Cielo device has been successfully connected to the Intern...

Page 34: ...ame is listed The device name can be renamed by the end user at their discretion The current system status is available on this page and provides information related to the readiness state of the syst...

Page 35: ...tant Make sure that the Cielo is successfully connected to the Internet To connect the Cielo Device to the internet please refer Section 2 3 Connect Azure Cielo Real Time PCR system to internet networ...

Page 36: ...m Port 587 Encryption STARTTLS MSN Server imap mail outlook com Port 993 Encryption SSL TLS Server pop mail outlook com Port 995 Encryption SSL TLS Server smtp mail outlook com Port 587 Encryption STA...

Page 37: ...ings Verify information has been entered correctly c Restart the device and try again 7 It is important follow these instructions in the order provided to enable the Email feature in the Protocol Setu...

Page 38: ...ny double stranded DNA fragments in an amplification reaction including nonspecific targets such as primer dimers By running a melt or dissociation curve at the completion of the run specificity of th...

Page 39: ...ding dye It will bind to any double stranded DNA molecule If there are primer dimer interactions or other off target primer to template binding SYBR Green will not distinguish between the desired targ...

Page 40: ...ndations on the best quencher to use with the fluorescent dye molecule chosen for detection While the probe is intact the quencher molecule is adjacent to the 5 end of the probe where the detection fl...

Page 41: ...on two different instruments the final data analysis results could be compared For example if an investigator was running a standard curve and then determining a copy number of a target that result c...

Page 42: ...ure Biosystems logo Azure Biosystems and Cielo are trademarks of the Company All other trademarks service marks and trade names appearing in this brochure are the property of their respective owners U...