3-20
Chapter 3. HOW TO USE THE LOADER
Adjusting the zero point
Note
•
To reset the zero point of a device with ID code “b” or later to the value at the
time when the device was shipped, set
[Zero point adjustment 1]
and
[Zero point
adjustment 2]
to “0.00” (the value at the time of shipment).
(
3 - 6 Immediate Writing of Parameters
•
If a device’s ID code is not indicated, it is not possible to restore the zero point
setting at the time when the device was shipped using the loader.
Instead, use the switch on the device to restore the setting.
For details, please refer to
Micro Flow Rate Liquid Flow Meter Model F7M User’s
Manual
(CP-SP-1421E).
•
If
[Automatic setting of flow rate correction factor (CCF)]
is enabled, the CCF
based on the amount of adjustment after zero point adjustment is automatically
set. If this function is used to automatically set the CCF, flow rate accuracy is not
guaranteed. For higher accuracy, use the set value as a guideline for setting the
CCF manually.
Handling Precautions
•
When zero point adjustment is in progress, do not turn off the device.
•
Parameters other than zero point adjustment are written to the device by
[Writing all parameters], but the amount of zero point adjustment is written to
the device when zero point adjustment is complete.
The procedure for adjusting the zero point of the instantaneous flow rate after
connecting the loader to the device is given below.
(1) Connect the loader to the device. (
3 - 2 Connecting the Loader to the
(2) Obtain device information by executing steps 1 to 4 in
parameters of the connected device (P. 3-8)
. If you have already executed
steps 1 to 4, this process is unnecessary.
(3) Select
[Device operation]
→
[Zero point adjustment]
on the menu bar.
>> The following screen is displayed.
3 - 8
Operating the Device