3-18
Chapter 3. HOW TO USE THE LOADER
3 - 7
List of Monitoring Parameters
[Monitor] tab
Parameters
Unit
Description
Factory
default
Process value Instantaneous
flow rate
mL/min Indicates the value calculated from the measured flow rate signal.
–
Abnormal status
–
Indicates the type of abnormal status.
–
Fluid settings
correction*
1
Instantaneous
flow rate
mL/min Indicates the value calculated from the measured flow rate signal.
–
Totalized flow
mL
Indicates the total volume of flow at that point in time.*
1
–
Totalized flow
count up
–
Totalizing of the flow volume can be started or stopped.*
1
*
2
Start
Stop
Stop and reset counter to 0
Stop
Liquid type
–
Indicates the currently selected liquid type. The type can be
changed.*
1
*
2
Liquid type 1
Liquid type 2
Liquid
type 1
Flow rate CCF 1
–
Indicates the flow rate correction factor for liquid type 1. The
value can be changed.*
1
*
3
0 to 100.000
1.000
Zero point
adjustment 1
mW
Indicates the amount of zero point adjustment for liquid type 1.
The value can be changed.*
1
*
3
-5.00 to +5.00
0.00
Flow rate CCF 2
–
Indicates the flow rate correction factor for liquid type 2. The
value can be changed.*
1
*
3
0 to 100.000
1.000
Zero point
adjustment 2
mW
Indicates the amount of zero point adjustment for liquid type 2.
The value can be changed.*
1
*
3
-5.00 to +5.00
0.00
Abnormal
status
Error
–
An error has a significant impact on the operation of the device.
Chapter 4. TROUBLESHOOTING (for details and corrective
actions)
–
Alarm
–
An alarm affects the measurement of flow rate.
Chapter 4. TROUBLESHOOTING (for details and corrective
actions)
–
Warning
–
A warning has a small impact on the measurement of flow rate.
Chapter 4. TROUBLESHOOTING (for details and corrective
actions)
–
Information
–
This is not technically an abnormal status. Information is
displayed for reference.
Chapter 4. TROUBLESHOOTING (for details and corrective
actions)
–
Sensor
information
Flow path
temperature
°C
Displays an approximate temperature in the flow path.
–
Circuit
temperature
°C
Displays the approximate temperature of the circuit.
–
Heater power
mW
Displays power consumption by the heater used for flow rate
measurement.
–