Troubleshooting
Issues Communicating with a Stimulator
Issue
Presentation
Resolution
CP cannot find Stimulator
Stimulator does not appear in list of
devices in the
Connect to Patient
Device
screen
•
Move closer to the Stimulator and hit the “Scan Again”
button on the screen to search again.
•
If the CP cannot find the Stimulator when close to the
implant site, verify with the patient’s Remote Control
(if one has been paired) that the Stimulator has battery
life and that there is not an error (red light on Remote
Control).
•
If the error light of the paired Remote Control is on or the
CP still cannot connect to the Stimulator, contact Axonics.
Communication with a
Stimulator is interrupted or lost
Error prompt appears stating that
communication with Stimulator has
been lost. If communication is not re-
established, CP will return to
Connect
to Patient Device
screen
•
Move closer to the Stimulator and hit the “Scan Again”
button on the screen to search again.
•
If the CP cannot find the Stimulator when close to the
implant site, verify with the patient’s Remote Control
(if one has been paired) that the Stimulator has battery
life and that there is not an error (red light on Remote
Control).
•
If the Remote Control error light is on or the CP still
cannot connect to the Stimulator, contact Axonics.
110-0054 rev E_output.indd 62
11/20/18 8:07 AM
Summary of Contents for SNM System 2501
Page 1: ...Sacral Neuromodulation System Model 2501 Clinician Programmer Manual ...
Page 2: ......
Page 4: ......
Page 5: ...1 Clinician Programmer Manual Model 2501 Clinician Programmer EN USA Rx ONLY ...
Page 8: ......
Page 71: ...67 Handbuch zum Programmiergerät Programmiergerät Modell 2501 DE USA Rx ONLY ...
Page 74: ......
Page 137: ...133 Manuel du programmateur du médecin Programmateur du médecin modèle 2501 FR USA Rx ONLY ...
Page 140: ......
Page 203: ...199 Manual del programador clínico Modelo 2501 Programador clínico ES USA Rx ONLY ...
Page 206: ......
Page 272: ......
Page 338: ......
Page 401: ......
Page 402: ......
Page 403: ......