47
Resetting the Stimulator
The Stimulator can be reset if there is a need to not deliver
stimulation for a period of time (e.g., days or longer) or if it was
configured with the wrong trial type. Resetting the EPG puts the
Stimulator in a power- conserving state to preserve the battery
of the Stimulator and erases all data from the Stimulator.
How to reset the Stimulator:
Press the “Reset” button ( 1 ).
A prompt will ask for confirmation before resetting the
Stimulator. The Stimulator will automatically disconnect from
the CP and the CP will return to the Home screen when the
Stimulator is reset.
Note:
The CP and patient Remote Control will not be able to communicate
with the Stimulator until the Stimulator is activated by pressing the button on
the back of the Stimulator.
How to wake up the Stimulator:
Press the button on the back
of the Stimulator. The green light will flash next to the button.
Connect with the CP within 90 seconds or the Stimulator will
turn off.
1
110-0054 rev E_output.indd 47
11/20/18 8:07 AM
Summary of Contents for SNM System 2501
Page 1: ...Sacral Neuromodulation System Model 2501 Clinician Programmer Manual ...
Page 2: ......
Page 4: ......
Page 5: ...1 Clinician Programmer Manual Model 2501 Clinician Programmer EN USA Rx ONLY ...
Page 8: ......
Page 71: ...67 Handbuch zum Programmiergerät Programmiergerät Modell 2501 DE USA Rx ONLY ...
Page 74: ......
Page 137: ...133 Manuel du programmateur du médecin Programmateur du médecin modèle 2501 FR USA Rx ONLY ...
Page 140: ......
Page 203: ...199 Manual del programador clínico Modelo 2501 Programador clínico ES USA Rx ONLY ...
Page 206: ......
Page 272: ......
Page 338: ......
Page 401: ......
Page 402: ......
Page 403: ......