![Axonics SNM System 2501 Programmer'S Manual Download Page 23](http://html.mh-extra.com/html/axonics/snm-system-2501/snm-system-2501_programmers-manual_3040489023.webp)
19
19
Understanding Impedance Values
Impedance values are important indicators when
delivering test stimulation or stimulating with the
Stimulator. The impedance values provided by the CP are
icons representing the quality of the circuit being used to
deliver stimulation.
Impedance can be tested when a
button is
present. When the button is pressed, the CP will display
either:
•
A single impedance icon corresponding to the
connection for foramen needle stimulation.
•
4 impedance icons, each corresponding to the
connection for one of the electrodes on the tined
lead.
The impedance icon can have one of 4 states:
1
Good
•
The impedance level indicates a
good connection and is preferred for
stimulation.
2
OK
•
The impedance level indicates an
acceptable connection and is allowed
for stimulation.
23(1
3
Bad – Open
•
The impedance level indicates an open
circuit and stimulation is not possible.
6+257
4
Bad – Short
•
The impedance level indicates a
short circuit and stimulation is not
recommended.
Note:
See the manual sections on Troubleshooting for steps to
resolve out of range (“Bad”) impedance values.
110-0054 rev E_output.indd 19
11/20/18 8:06 AM
Summary of Contents for SNM System 2501
Page 1: ...Sacral Neuromodulation System Model 2501 Clinician Programmer Manual ...
Page 2: ......
Page 4: ......
Page 5: ...1 Clinician Programmer Manual Model 2501 Clinician Programmer EN USA Rx ONLY ...
Page 8: ......
Page 71: ...67 Handbuch zum Programmiergerät Programmiergerät Modell 2501 DE USA Rx ONLY ...
Page 74: ......
Page 137: ...133 Manuel du programmateur du médecin Programmateur du médecin modèle 2501 FR USA Rx ONLY ...
Page 140: ......
Page 203: ...199 Manual del programador clínico Modelo 2501 Programador clínico ES USA Rx ONLY ...
Page 206: ......
Page 272: ......
Page 338: ......
Page 401: ......
Page 402: ......
Page 403: ......