Viewing the Status of a Stimulator
Stimulator Errors
If a Stimulator error is present, a “Query Fault” button will
appear to the right of the “Disconnect” button in the lower right
corner of the
Patient Device
screen.
When the “Query Fault” button is pressed, a pop-up will appear
( 1 ) with a description of the error ( 2 ).
To resume the communication session and try to resolve the
error with reprogramming, press “Clear Error” ( 3 ) to remove
the error notification from the Stimulator. If additional errors
exist, the pop-up will
display information about the next error, which will also need to
be cleared.
Press “Close” to exit the pop-up ( 4 ). If the pop-up is exited
without clearing the errors, the ability to program the Stimulator
may be impaired.
1
2
3
4
110-0054 rev E_output.indd 45
11/20/18 8:07 AM
Summary of Contents for SNM System 2501
Page 1: ...Sacral Neuromodulation System Model 2501 Clinician Programmer Manual ...
Page 2: ......
Page 4: ......
Page 5: ...1 Clinician Programmer Manual Model 2501 Clinician Programmer EN USA Rx ONLY ...
Page 8: ......
Page 71: ...67 Handbuch zum Programmiergerät Programmiergerät Modell 2501 DE USA Rx ONLY ...
Page 74: ......
Page 137: ...133 Manuel du programmateur du médecin Programmateur du médecin modèle 2501 FR USA Rx ONLY ...
Page 140: ......
Page 203: ...199 Manual del programador clínico Modelo 2501 Programador clínico ES USA Rx ONLY ...
Page 206: ......
Page 272: ......
Page 338: ......
Page 401: ......
Page 402: ......
Page 403: ......