Setting Up a New Stimulator
Checking the Lead Connection
An icon showing the impedance status of each electrode
is shown to the right of the graphic representation of each
electrode ( 1 ). See manual section
Understanding Impedance
Values
for more information about the qualitative impedance
indicators.
The CP automatically checks impedance when it first connects
to the Stimulator. To check the impedances again, press the
impedance button:
( 2 ).
Impedances may change from green to yellow as fibrous tissue
encapsulates the electrodes. This change will not affect a
patient’s therapy because the current-controlled Stimulator
automatically compensates for impedance changes.
Impedance changes from green or yellow to red may indicate an
open circuit, which may be caused by a disconnected cable or
other problem with the device.
Note:
If any impedances are red during Stimulator set up, check the
connections between the lead and the Stimulator.
1
2
110-0054 rev E_output.indd 39
11/20/18 8:07 AM
Summary of Contents for SNM System 2501
Page 1: ...Sacral Neuromodulation System Model 2501 Clinician Programmer Manual ...
Page 2: ......
Page 4: ......
Page 5: ...1 Clinician Programmer Manual Model 2501 Clinician Programmer EN USA Rx ONLY ...
Page 8: ......
Page 71: ...67 Handbuch zum Programmiergerät Programmiergerät Modell 2501 DE USA Rx ONLY ...
Page 74: ......
Page 137: ...133 Manuel du programmateur du médecin Programmateur du médecin modèle 2501 FR USA Rx ONLY ...
Page 140: ......
Page 203: ...199 Manual del programador clínico Modelo 2501 Programador clínico ES USA Rx ONLY ...
Page 206: ......
Page 272: ......
Page 338: ......
Page 401: ......
Page 402: ......
Page 403: ......