SAFETY CONCERNS
TURN OUTPUT OFF BEFORE ACCESSING DUT
Always turn the output off before opening the rear-panel DUT door, using the front
panel menu or by computer command (“output off”).
If you do not disable the output first, the output will be automatically disabled by a
sensor on the DUT door when it is opened. However, the user should not rely on this
for safety purposes.
Always check that the front panel indicates that the output is “off” before touching the
DUT area.
For maximum safety, turn off the instrument before accessing the DUT area. Ensure
that the instrument has been turned off for at least 60 seconds before turning it back on,
to allow the embedded controller to reset properly.
AUTOMATIC TIMEOUT
The output is disabled automatically after 90 seconds of inactivity. The timer resets
each time a pulse parameter (amplitude, PRF, etc) is changed.
OVERLOAD INDICATOR
The front-panel “overload” indicator should always glow green, and never amber
(except briefly when the instrument is first turned on).
If the overload indicator becomes active (amber), stop using the instrument
immediately.