d) Press CHANGE to return to the main menu.
5. To set the HV PULSE amplitude to +1.5 kV:
a) Press the MOVE button until the arrow pointer is pointing at the AMP1 menu
item.
b) Press the CHANGE button. The submenu will appear. Rotate the ADJUST
knob until the amplitude is set at +1.5 kV.
c) Press CHANGE to return to the main menu.
6. To set the VCC2 amplitude to +5V:
a) Press the MOVE button until the arrow pointer is pointing at the AMP2 menu
item.
b) Press the CHANGE button. The submenu will appear. Rotate the ADJUST
knob until the amplitude is set at +5V.
c) Press CHANGE to return to the main menu.
7. At this point, nothing should appear on the oscilloscope.
8. To enable the output:
a) Press the MOVE button until the arrow pointer is pointing at the output menu
item.
b) Press the CHANGE button. The output submenu will appear.
c) Press MOVE until the arrow pointer is pointing at the “ON” choice.
d) Observe the oscilloscope. You should see +1.5 kV pulses with < 10 ns rise
time, approximately.
e) Press CHANGE to return to the main menu.
9. 90 seconds after the last parameter change, the instrument will automatically shut
the output off. Confirm that this occurs.
This completes the operational check.