Performing a measurement
AVL DiTEST HV Safety 2000
3-14
User manual
3.8
Capacity measurement
In the
"Capacity measurement"
mode capacitors can be measured in the range 1 nF to 300
μF.
Please note that the component which is to be assessed must be disconnected and that all cable
connections which might affect the measurement must be removed prior to each capacity
measurement.
1.
When you click on the
Capacity measurement
mode (see Fig. 3-1), a screen of safety
instructions will appear.
You must read and observe these safety instructions.
2.
Click on
F8 Next
.
3.
The AVL DiTEST HV Safety 2000
safety test
is carried out:
- Connect the probe tips briefly
- Press and hold the test button on the red test adapter until the safety test has completed
- The result of the safety test is displayed
4.
Place the probe tips on the test object.
5.
Press and hold the test button on the red test adapter until a stable measured value is
displayed.
6.
The result of the measurement is displayed, see Fig. 3-8
(If the test button is not kept pressed for long enough, the following message will appear:
“
Measurement invalid: Test button released too early
”. If this happens, start a new
measurement.
Press and hold the test button on the red test adapter until a stable measured value is
displayed.
Fig. 3-8
DANGER
Summary of Contents for DITEST HV SAFETY 2000
Page 18: ...General AVL DiTEST HV Safety 2000 1 8 User manual ...
Page 44: ...Troubleshooting AVL DiTEST HV Safety 2000 4 4 User manual ...
Page 46: ...Maintenance and care AVL DiTEST HV Safety 2000 5 2 User manual ...
Page 48: ...Warranty AVL DiTEST HV Safety 2000 6 2 User manual ...
Page 56: ...Technical data AVL DiTEST HV Safety 2000 7 8 User manual ...
Page 58: ...Index AVL DiTEST HV Safety 2000 8 2 User manual ...