extension will be *.fbr. The name of the intensity calibration file can be entered after clicking
the “Save As” button.
10.
Switch to the Irradiance Chart TAB to enter the hardware setup and select the colorimetric,
radiometric, photometric and/or peak parameters of interest. Then click OK.
11.
Measure the output parameters in the experiment. If needed, change the integration time, such
that the maximum in Scope Mode is around 14000 A/D Counts . Block the light path to the
spectrometer, and save a dark spectrum. If the (ir)radiance of the light to be measured needs to
be displayed against time, click the time measurement TAB in the settings dialog as described in
section 4.4.6.
12.
The intensity calibration as performed under point 9 can be loaded in future experiments by
selecting the option “Load Intensity Calibration”, as described below under Quick Start (2).
After loading an intensity calibration, a dark spectrum needs to be saved before switching to
Irradiance mode.
Technical Data
Wavelength Range
350-1095 / 1100-2500nm
Calibration Accuracy
Repeatability
± 0.5 %
Calibration Relative
Uncertainty to NIST standard
± 9.5 % (350-1100nm)
± 6.5% (1100-1950nm)
± 9,5% (1950-2500nm)
Calibration valid for
60 hrs
Time to stabilize
Ca. 15 Min.
Bulb Output (CC-UV/VIS)
Ca. 170µW/cm
2
nm (@800nm)
Power requirement
24 VDC / 1.2A
Dimensions (mm)
132 x 110 x 44
Ordering information
AvaLight-HAL-CAL
NIST traceable Halogen Lamp with CC-UV/VIS-CAL diffuser, incl. PS-
24V/1.25A
AvaLight-HAL-CAL-
ISP30
NIST traceable Halogen Lamp for use with AvaSphere-30-IRRAD, incl. PS-
24V/1.25A
AvaLight-HAL-CAL-
ISP50
NIST traceable Halogen Lamp for use with AvaSphere-50-IRRAD, incl. PS-
24V/1.25A
AvaLight-HAL-CAL-
ISP80
NIST traceable Halogen Lamp for use with AvaSphere-80-IRRAD, incl. PS-
24V/1.25A
HL-Recal
AvaLight-HAL-CAL recalibration service 350-1095nm
HL-Recal-NIR
AvaLight-HAL-CAL extended or recalibration service 1100-2500nm