102
Apr-09
Avantes
www.avantes.com
the intensity calibration data will be saved to an ASCII file with extension *.dfr, with bare
fiber this extension will be *.fbr. The name of the intensity calibration file can be entered
after clicking the “Save As” button.
10.
Switch to the Irradiance Chart TAB to enter the hardware setup and select the colorimetric,
radiometric, photometric and/or peak parameters of interest. Then click OK.
11.
Measure the output parameters in the experiment. If needed, change the integration time,
such that the maximum in Scope Mode is around 14000 A/D Counts . Block the light path to
the spectrometer, and save a dark spectrum. If the (ir)radiance of the light to be measured
needs to be displayed against time, click the time measurement TAB in the settings dialog.
12.
The intensity calibration as performed under point 9 can be loaded in future experiments by
selecting the option “Load Intensity Calibration”, as described below under Quick Start (2).
After loading an intensity calibration, a dark spectrum needs to be saved before switching to
Irradiance mode.
Technical Data
Wavelength Range
200-1099 / 1100-2500nm
Calibration Accuracy
Repeatability
± 1 %
Calibration Relative
Uncertainty to NIST standard
± 10 % (200-400nm)
± 9 % (240-350nm)
± 10% (350-400nm)
± 9.5% (400-1100nm)
Calibration valid for
60 hrs
Time to stabilize
Ca. 30 Min.
Bulb Output (CC-UV/VIS)
Ca. 80µW/cm
2
nm (@215nm)
Ca. 5µW/cm
2
nm (@800nm)
Power requirement
100-240VAC
Dimensions (mm)
315x165x140
Ordering Information
AvaLight-DH-CAL
NIST traceable UV/VIS Deuterium/Halogen Lamp with CC-UV/VIS-CAL
diffuser, -SR fibers recommended
AvaLight-DH-CAL-ISP30
NIST traceable UV/VIS Deuterium/Halogen Lamp for use with AvaSphere-30-
IRRAD, -SR fibers recommended
AvaLight-DH-CAL-ISP50
NIST traceable UV/VIS Deuterium/Halogen Lamp for use with AvaSphere-50-
IRRAD, -SR fibers recommended
AvaLight-DH-CAL-ISP80
NIST traceable UV/VIS Deuterium/Halogen Lamp for use with AvaSphere-80-
IRRAD, -SR fibers recommended
DH-Recal
AvaLight-DH-CAL recalibration service 200-1099nm
DH-Recal-NIR
AvaLight-DH-CAL extended or recalibration service 1100-2500nm