26
Apr-09
Avantes
www.avantes.com
info@avantes.com
Resolution table (FWHM) for AvaSpec-2048
*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the higher
the resolution
Ordering Information
AvaSpec-2048
Fiber Optic Spectrometer, 75 mm Avabench, 2048 pixel CCD detector,
USB1.1/RS-232 interface, incl AvaSoft-Basic, USB cable and PS-12V/1.0A
power supply ,specify grating, wavelength range and options
AvaSpec-2048-USB2
Fiber Optic Spectrometer, 75 mm Avabench, 2048 pixel CCD detector, USB
powered high speed USB2 interface, incl. AvaSoft-Basic, USB interface cable,
specify grating, wavelength range and options
AvaSpec-2048-2
Dual channel Fiber Optic Spectrometer, 2 * 75 mm Avabench, 2048 pixel CCD
detector, USB1.1/RS-232 interface, incl AvaSoft-Basic, USB cable and PS-
12V/1.0A power supply ,for both channels specify grating, wavelength range
and options
Options
-SPU
incl. switch for self powered USB1 or external power for RS232
-SPU2
incl. switch for self powered USB2 or external power for RS232
-SPU2-BT
Bluetooth interface for USB2 platform only, including antenna and switch
SDXXX
Internal XXX MB SD card for on board data saving, for USB2 platform only
DUV
Deep UV detector coating >150 nm
DCL-UV/VIS
Detector Collection lens to enhance sensitivity, Quartz, 200-1100 nm
SLIT-XX
Slit size, please specify XX = 10, 25, 50, 100, 200, 500 µm
OSF-YYY
Order sorting filter for 2
nd
order effects filtering, please specify YY= 375,
475, 515, 550, 600 nm
OSC
Order sorting coating with 590nm long pass filter for VA, BB (>350nm) and
VB gratings in AvaSpec-2048
OSC-UA
Order sorting coating with 350 and 590nm longpass filter for UA gratings in
AvaSpec-2048
OSC-UB
Order sorting coating with 350 and 590nm longpass filter for UB or BB
(<350nm) gratings in AvaSpec-2048
Purge
set of 4 mm Gasfittings for AvaSpec purging in DUV (<200nm) applications
N.B. In order to change a grating, wavelength range or any of the options, the unit needs to be
returned to Avantes manufacturing, please ask for an RFU number (see page 4 of this manual).
The cost for the socalled AvaSpec-Upgrade depends on the modification that needs to be done.
Slit size (µm)
Grating
(lines/mm)
10
25
50
100
200
500
300
0.8
1.4
2.4
4.3
8.0
20.0
600
0.4
0.7
1.2
2.1
4.1
10.0
1200
0.1-0.2*
0.2-0.3*
0.4-0.6*
0.7-1.0*
1.4-2.0*
3.3-4.8*
1800
0.07-0.12* 0.12-0.21*
0.2-0.36*
0.4-0.7*
0.7-1.4*
1.7-3.3*
2400
0.05-0.09* 0.08-0.15* 0.14-0.25*
0.3-0.5*
0.5-0.9*
1.2-2.2*
3600
0.04-0.06* 0.07-0.10* 0.11-0.16*
0.2-0.3*
0.4-0.6*
0.9-1.4*