0611
Verifit User’s Guide Version 2.8
Page 92
19.5 Fitting to targets for soft speech
FastFacts 19.5: Fitting to targets for soft speech
See
18.9: REM or S-REM mode
,
18.6: Using Speechmap
and
19.2: Assessment
data entry
for setup and audiometric data entry.
1)
Select [Audiometry] to choose the target method and enter audiometric data.
2)
Select REAR 1 and set stimulus to [Speech-std(1)] or [Speech-std(2)] and level
to 50. Elongated
+
symbols show the target range for 50 dB speech LTASS.
3)
Adjust frequency shaping and gain for soft sounds so that the middle curve (the
LTASS) falls within the target range, especially between 500 - 4000 Hz. Press
<Continue> to run the complete passage. Repeat as necessary.