17.3 Insertion gain in HL
FastFacts 17.3: Insertion gain in HL
See
10: Real-Ear Measurement Setup
and
11: REM Screen Setup
1)
Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list.
2)
Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body.
This changes the CORFIG used for 2cc targets.
3)
Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R,
POGO II, BERGER, LIBBY, USER, NONE.
An estimated aided threshold curve is calculated by shifting the unaided threshold
curve upward by the REIG. This estimate assumes
a) the
unaided threshold is a
sound-field threshold
and b) a
linear aid
(i.e. REIG doesn’t depend on level).
The REUR curve uses the right-hand SPL scale and is shown for reference only.