Page 50
9.4
Directional ITE positioning
FastFacts 9.4: Directional ITE positioning
Attach the provided clip to the coupler for stability. Orient the instrument as shown
so that the direction of maximum sensitivity is toward one of the speakers and a
line joining the microphone ports is at 45° and horizontal. Locate the HIT reference
microphone at the front microphone opening on the faceplate and seal any vent at
the faceplate. Note which speaker (L or R) is in “front” of the hearing instrument for
reference. Other positions may be preferable for some instruments.