MAN - BBR3 - REV: Original | E. Operation
27
the Setup screen (Figure E.4).
After you set your temperature, press “Main Menu” to return to the main BBR3 screen. From the Main Screen press
the “Standardization” button to enter the Standardization Screen (Figure E.5).
NOTE: The standardization button will not be shown on the main menu unless the currently logged in
user has permission to perform standardization.
The BBR3’s system status lights, located at the bottom right of the screen, indicate whether or not a component
requires standardization.
• Red = the component requires standardization
•
Yellow = the component requires verification
• Green = the component is ready for testing
Prior to standardization, verify that the information entered into the necessary fields matches the information provided
on the Certificate of Conformance that was shipped with the gauge kit being used for the standardization. Make
adjustments as necessary. You will also need a BBR3 gauge kit.
Figure E.5 - Standardization Screen
Figure E.4 - “Bath Temperature” Field on Setup Screen