45
Operating instructions XENOTEST BETA+
7.
Operation
5. Remove the XenoCal BB 300-400 sensor from the test chamber and read the value of
the measured irradiation intensity.
6. Select the Settings program module:
Press the
p
key
7. Select the following program functions in succession:
• Press the
A
djustment
switch,
• Press the
A
djustment
E
switch.
8. Enter the irradiation intensity value read with the XenoCal BB 300-400 sensor into
the input field. The display now shows the old adjustment factor and the newly
calculated adjustment factor.
If the new factor is not within the tolerance (below 0.8 or above 1.2), send the
XENOSENSIV sensor in to ATLAS for a check.
9. Confirm the new adjustment factor:
Press the
b
key
After an adjustment has been made, the adjusted test parameters should be calibrated
(measured value check).
Adjusting the black standard temperature:
1. Interrupt the test:
Press the
n
key
2. Insert the operational XenoCal BST black standard temperature sensor instead of
a sample holder into the sample holder rack.
3. Resume the test (without spraying, dark phase or sample backside cooling):
Press the
t
key
4. After approx 45 minutes (the run time should be at least 45 minutes), interrupt the test.
Press the
n
key
5. Remove the XenoCal BST sensor from the test chamber and read the value of the
measured temperature.
6. Select the Settings menu:
Press the
p
key
7. Select the following program functions in succession:
• Press the
A
djustment switch,
• Press the
A
djustment
BST
switch.