46
Operating instructions XENOTEST BETA+
7.
Operation
8. Enter the black standard temperature value read with the
XenoCal BST sensor into the input field. The display now
shows the old adjustment factor and the newly calculated
adjustment factor.
If the new factor is not within the tolerance (below 0.8 or
above 1.2), send the XENOSENSIV sensor in to ATLAS
for a check.
9. Confirm the new adjustment factor:
Press the
b
key
After an adjustment has been made, the adjusted test parameters should be calibrated
(measured value check).
7.6
Control without senors
Controlling the irradiation intensity without XENOSENSIV sensor:
If the device is operated without the XENOSENSIV sensor, the control is perfomed in
accordance with the relative rating of the Xenon lamp within a range of 53 to 120 %.
100 % lamp rating correspond with the nominal rating of 2200 VA.
The control range of 100 to 120 % considers lamp aging.
NOTE - Operation without sensor
If the XENOTEST BETA+ is operated without
XENOSENSIV sensor, the adapter for opera-
tion without sensor (terminating plug) must be
connected to the connecting sleeve (see section
6.6).