www.atestgaz.pl
Bit
Flag
Description
9
Test
Test mode
10
Warm_Up
Sensor's warm up
11
Sensor_Inhibit
Inhibit mode
12
Comm_Error
Error of communication with a detector
13
Calibration_Warning
Calibration time exceeded (non-critical error)
14
Monitoring
Measurement in progress
15
System_Stop
The system is stopped
N – gas concentration. The value of 0 corresponds to zero concentration whilst the value of 1000
corresponds to the concentration equal to the measurement range of the detector.
1.2
Status of R1 – R8 outputs and DI1 – DI4 inputs (read-only)
Register
Name
Description
Type / range
40065
DO_Status
Status of power supply for solenoids of R1 – R8 relays
Subsequent bits correspond to individual outputs:
bit 0 – R1; …; bit 7 – R8
Bit value of 1 (high): the R relay is active
Bit value of 0 (low): the R relay is inactive
flags
40066
DI_Status
Status of DI1 inputs
Subsequent bits correspond to individual inputs:
bit 0 – DI1; …; bit 3 – DI4
Bit value of 1 (high): the DI input is active
Bit value of 0 (low): the DI input is inactive
flags
1.3
Status of the control unit (read-only)
Register
Name
Description
Type / range
40067
CU_Status
Status of the Control Unit
flags
CU_Status – status of control unit. Assignment of the individual bits is listed in the table below.
Bit
Flag
Description
0
System_fail
Collective flag of system failure
1
CU_fail
Failure of the control unit
2..15
-
Unused
Atest-Gaz
A. M. Pachole sp. j.
ul. Spokojna 3, 44-109 Gliwice, Poland
VAT No.: PL 9691433231
tel.: +48 32 238 87 94
fax: +48 32 234 92 71
e-mail: [email protected]
Appendix: PU-Z-006-ENG R03
p. 2|4
Summary of Contents for PW-033-B
Page 1: ...POD 003 ENG R04 User Manual Control Unit Module Sigma MOD LCD Product code PW 033 B...
Page 49: ...www atestgaz pl Notes...
Page 50: ...www atestgaz pl Notes...
Page 51: ......