www.atestgaz.pl
internal buzzer is activated, however it can be muted for a specific time by means of a keypad pushbutton
(see Section 3.1.5).
When gas concentration exceeds the maximum range permissible for the detector (overload threshold), the
following actions take place:
for detectors with catalytic sensors – the detector is locked (its sensor is deactivated and the
detector latches the most recent measurement result). The corresponding LED in the area for
status indication goes on (see Section 3.3.2). To restore normal operation of the detector the
detector must be unlocked from another device connected to the system,
for detectors with other sensors – the corresponding LED in the status area, depicted as . goes
on. The detector keeps working and continues measurements of gas concentration, which is
confirmed by the MEAS. indicator that is still on.
Signalling depending on the concentration of gas measured by the detector:
s. 23|31
User Manual: POD-002-ENG R03
Figure 14: Indications of LEDs on the front panel versus gas concentration measured by the detector
ALARM
2
1
MEAS.
Warning 2
Gas
concentration
Warning 1
ALARM
Overload
ALARM
2
1
MEAS.
S
e
ns
or
S
n
0
2
,
P
ID
e
le
ct
ro
ch
em
ic
a
l,
IR
C
at
al
yt
ic
s
e
ns
or
N
o
da
ng
er
Description
N
o
d
an
ge
r
time
W
ar
n
in
g
1
W
a
rn
in
g
1
W
a
rn
in
g
2
W
ar
n
in
g
1
W
a
rn
in
g
2
A
la
rm
Sensor lock –
sensor off, the last
measurement latched
The gas concentration is
falling, and with it
disappears warnings
W
ar
m
-
up
O
p
er
a
tio
n
"L
oc
k
ca
nc
e
l"
.
Gas
overload
Summary of Contents for PW-033-A
Page 40: ...www atestgaz pl Notes...
Page 41: ...www atestgaz pl Notes...
Page 42: ...www atestgaz pl Notes...
Page 43: ......