www.atestgaz.pl
1.6
Interface for execution of operator commands (read /write; available only for Sigma MOD LCD)
Register
Name
Description
Type /range
43501
Command_Status
Current status of the command execution
Write – data byte is ignored
Read – status, the available values are:
•
0 – inactive
•
1 – command in progress
•
2 – command successfully completed; the value is
maintained for 5s after completion of the command
execution
•
3 – command execution failed, invalid command or its
parameters; the value is maintained for 5s after completion
of the command execution
U16
43502
Command_Code
Code of the command to be executed
Write – code of the command to be executed, loading to this
register initiates execution of the command
Read – current value of the command code
Available values – see the description below
U16
43503
Command_Param
Command parameter
Read, write – command parameter
Available values – see the description below
U16
List of commands:
Command code
Description
1
Detector's lock reset
Execution parameters – no. of the sensor's channel where the sensor lock is to be cleared,
available values: 1 – 32
Atest-Gaz
A. M. Pachole sp. j.
ul. Spokojna 3, 44-109 Gliwice, Poland
VAT No.: PL 9691433231
tel.: +48 32 238 87 94
fax: +48 32 234 92 71
e-mail: [email protected]
Appendix: PU-Z-006-ENG R03
p. 4|4
Summary of Contents for PW-033-A