EF600 Instruction Manual
Asymptote Ltd.
St John’s Innovation Centre, Cowley Road, Cambridge, CB4 0WS, UK
Tel: +44 1223 421161, Fax: +44 1223 421166, E-mail: [email protected]
14
correctly.
Figure 5.4 – Straws positioned in sample plate
5.5
Starting the data logging (EF600-01, EF600-10)
To start the data logging, click on the grey ‘START LOGGING’ button on the
user interface. A message box will appear, prompting the user to check that an
appropriate file name has been given. To go back and check, press ‘Cancel’;
to start logging, click on ‘OK’.
5.6
Starting the freezing cycle
When the straws have been positioned correctly, wait for at least one minute
before starting the freezing cycle to allow the straws to reach thermal
equilibrium with the plate. Note that the freezer must also have been switched
on for at least two minutes in order for the controller to ‘initialise’.
Start the freezing cycle by pressing the ‘Start Cycle’ button on the keypad.
The ‘Cycle On’ button will illuminate green and will remain on for the entire
freezing cycle.
Note that the ‘Ready To Load’ light may remain on for a short time at the