CLF Probe calibration
Do not calibrate the probe until
the water in the pool is thoroughly mixed
and
the value on the ASIN AQUA Net display is stable.
This may take several hours.
Calibration is performed by entering the manually measured value of chlo-
rine concentration (using a photometer) in the CALIBRATION menu.
Calibration
is not necessary
if the difference between the photometer mea-
sured value and the value shown on the display
is less than 0.2 mg / liter
.
Calibration is best performed with chlorine concentrations in the pool water
in the range of
0.3 - 1.2 mg / l
.
It is best to calibrate to a value equal to or greater than the desired value.
0,3
0,4
0,4
0,7
CHLORINE
pH
ALGICID
FLOCC
0,5
0,3 – 1,2 mg/l
Photometr
# 13076
Calibration restrictions
The CLF probe cannot be calibrated if the output
signal is less than 20 mV
.
The CLF probe can only be calibrated in the CL range
from 0.3 to 5.0 mg / l
.