Preventative Maintenance Inspection Form
DFU-0276-4r0_fmt_en-US
PM – 6/8
Electrical Safety Test:
Record the Model number, Serial number, and Calibration date of the Electrical Safety Analyzer.
Model Number: __________________________ Serial number: _____________________
Calibration Date: __________________________
Perform the following Electrical Safety Tests
Measurement
Measured Value
Limit /
Test Level
Pass/Fail
Operating Voltage
N/A
Operating Current
N/A
Ground Impedance Test
(Measure from Ground on inlet terminal to
metal chassis)
<0.1 ohm
Leakage Current
(Chassis)
Normal Condition
<0.1mA
Leakage Current
(Chassis)
Single Fault –
<0.5mA
Earth Leakage Current
Normal Condition
<0.5mA
Earth Leakage Current
Single Fault –
<1.0mA
*Test limits provided are per IEC 60601-1 3
rd
Edition and IEC 62353:2014-09
Results
If all steps have passed or measured within tolerance then the device can be put back into service.
If an item has failed please contact Arthrex Technical Assistance Center at Toll-Free Technical Support:
1-(888) 420-9393, Monday through Friday, 9:00 AM – 5:00 PM EST.
Summary of Contents for SynergyRF AR-9800
Page 2: ...Page i This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Page 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Page 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Page 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Page 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...