Product Drawings and References
Synergy
RF
™
System
Service Guide
DFU-0276-4r0_fmt_en-US
Page 15 of 50
2.4.3
Apollo
RF
Service probe
Figure 7
Apollo
RF
Service probe - Description
Table 8
Apollo
RF
Service probe - Description
1.
Ablation Power
Change
button
Each time the button is pressed, the Ablation
setting increases by 1 unit. Activation increases
ablation in a positive direction only. When the
device is at maximum power setting, the user can
return to the minimum setting by pressing the
Ablation Power Change
button one more time.
2.
Ablate
button
Initiates ablation current.
3.
Coagulation
button
Initiates coagulation current.
4.
Red (+) output
connector
Connects to the red variable load port of
electrosurgical analyzer.
5.
Black (-) output
connector
Connects to the black variable load port of
electrosurgical analyzer.
6.
Connector Cable Cord
Connects the Service probe to the Synergy
RF
console.
The Apollo
RF
Service probe is designed to be used by trained biomedical technicians
ONLY
.
The device initiates the electrical current from the Synergy
RF
console, through the service
probe, and into the electrosurgical unit (
ESU
) analyzer.
It is intended for ESU analysis only.
It is NOT intended for use in surgical applications or to be operated by surgeons or
surgical staff.
1
2
3
4
5
6
Summary of Contents for SynergyRF AR-9800
Page 2: ...Page i This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Page 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Page 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Page 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Page 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...