Synergy
RF™
System
Service Guide
DFU-0276-4r0_fmt_en-US
Page 35 of 50
Table 17
Guidance and Manufacturer's Statement - Electromagnetic Immunity
The AR-9800 Synergy
RF
System is intended for use in the electromagnetic environment specified
below. The customer or the user of the AR-9800 Synergy
RF
System should ensure that it is used
in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
100khz Cycling
Frequency
± 2 kV for power
supply lines
± 1 kV for input/output
lines
100khz Cycling
Frequency
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Uτ
= 0%, 0.5 cycle
(0,45,90,135,180,225,270
and 315º)
Uτ
= 0%, 1 cycle
Uτ
= 70%; 25/30 cycles
(@ 0 degrees)
Uτ
= 0%; 250/300 cycles
(
Uτ
= 0%, 0.5 cycle
(0,45,90,135,180,225,270
and 315º)
Uτ
= 0%, 1 cycle
Uτ
= 70%; 25/30 cycles
(@ 0 degrees)
Uτ
= 0%; 250/300 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the AR-9800 Synergy
RF
System requires
continued operation
during power mains
interruptions, it is
recommended that the
AR-9800 Synergy
RF
System be powered from
an uninterruptible power
supply.
Summary of Contents for SynergyRF AR-9800
Page 2: ...Page i This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Page 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Page 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Page 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Page 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...