Synergy
RF™
System
Service Guide
DFU-0276-4r0_fmt_en-US
Page 39 of 50
Table 18
Guidance and Manufacturer's Statement – Recommend Separation
Recommended separation distances between portable and mobile RF communications
equipment and the Model
AR-9800
The Model
AR-9800 Synergy
RF
System is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Model AR-9800
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Model AR-9800 as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter [
W]
Separation distance according to frequency of transmitter [
m]
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
150 kHz to 80 MHz
d = 1.2 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for SynergyRF AR-9800
Page 2: ...Page i This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 20 of 50 ...
Page 27: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 21 of 50 ...
Page 28: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 22 of 50 ...
Page 29: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 23 of 50 ...
Page 30: ...Technical Specifications SynergyRF System Service Guide DFU 0276 4r0_fmt_en US Page 24 of 50 ...