INSTRUCTIONS FOR USE
Document Number: D-720763
Page 29
Issue Date: 31 MAR 2020
Version: C
Ref Blank Template: 80025118 Ver. E
Patient Platform:
Disengage the Table Clamp lock by pulling the silver lock button distally with your
thumb while pivoting the Patient Platform superiorly. Pull the Patient Platform away
from the OR table so that the Patient Platform forks disengage from the Table
Clamps. Clean as directed in the following section and store in a safe location,
preferably the designated location on the HDS Storage Cart.
Table Clamps:
Loosen the knob on the lateral aspect of the Clamp body to unlock the Clamp from
the table rail. Slide the Table Clamp off the rail, disengaging the table rail gravity
locks if appropriate. Clean as directed in the following section and store in a safe
location, preferably the designated location on the HDS Storage Cart.
3.5
Troubleshooting Guide:
This device does not have a troubleshooting guide. For technical support users of
the device shall first contact their supplier.
3.6
Device Maintenance:
Make sure that all labels are installed and can be read. Replace labels as
necessary by using a plastic scraper to remove the label. Use an alcohol wipe to
remove any adhesive residue.
Contact us if you need to repair or replace the device by using the information
from the contact details section (1.3).
4
Safety Precautions and General Information:
4.1
General Safety Warnings and Cautions:
WARNING:
a.
Do not use if product shows visible damage.
b.
Prior to using this device, please read the instructions for equipment set up and
use. Familiarize yourself with the product before application on a patient.
c.
To prevent patient and/or user injury and/or equipment damage, examine the
device and surgical-table side rails for potential damage or wear prior to use. Do
not use the device if damage is visible, if parts are missing or if it does not function
as expected.
d.
To prevent patient and/or user injury and/or equipment damage, verify the device
attaching clamps completely touch the table-side rails and are firmly in place. Test
the locking mechanism to ensure no movement when elevated or pushed.
Summary of Contents for AR-6529S
Page 1: ...Hip Distraction System Instructions for Use Product No AR 6529S D 720763 Version C ...
Page 34: ...Hip Distraction System Instructions d utilisation N de produit AR 6529S D 720763 Version C ...
Page 67: ...Hip Distraction System Gebrauchsanleitung Produkt Nr AR 6529S D 720763 Version C ...
Page 101: ...Hip Distraction System Istruzioni per l uso N prodotto AR 6529S D 720763 Version C ...
Page 135: ...Hip Distraction System Instrucciones de uso N º de producto AR 6529S D 720763 Version C ...
Page 173: ......
Page 174: ......