DFU-0332r0_fmt_en-US
Page 45 of 64
3.3
Cleaning and Sterilizing
Follow universal precautions for protective apparel when handling and cleaning contaminated
instruments.
3.3.1
Cleaning the LED Light Engine
1. Allow LED Light Engine to cool for ½ hour before cleaning.
2. Dampen one end of a cotton swab with isopropyl alcohol.
3. Clean any residue from optic using circular motion.
4. Use the DRY END of the cotton swab to dry the face of the optics.
5. Inspect the optics for residue or cotton fibers and clean as required.
6. Allow to AIR DRY for 5 minutes prior to use.
3.3.2
Cleaning the Camera Control Unit (CCU)
1. Turn the CCU power off. Disconnect the power cord from the electrical power source and from
the rear of the CCU.
2. Remove the camera head from the CCU.
3. Use a fresh CaviWipe or equivalent to remove any gross contamination from the device.
4. Use a second fresh CaviWipe or equivalent to thoroughly wet the surface of the device and
ensure that it remains visibly wet for 3 minutes. Additional CaviWipes or equivalents may be
needed to ensure that the device remains visibly wet for 3 minutes.
5. Allow the device to air dry.
3.3.3
Cleaning the Tablet
1. Disconnect the Tablet from power.
2. Wipe the Tablet Touchscreen with a clean, soft cloth dampened with a mild pH-balanced
detergent.
CAUTION: Do not sterilize the CCU/Tablet or immerse in liquid or disinfectant solution.