DFU-0332r0_fmt_en-US
Page 59 of 64
IEC 60601-1-2
Guidance and manufacturer’s declaration – electromagnetic emissions
The ArthrexID Video System is intended for use in the electromagnetic environment specified below. The customer or
the user of the ArthrexID
Video System should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment- guidance
RE emissions
CISPR 11
Group 1
The ArthrexID Video System uses RF energy only for its internal
function. Therefore its RF emissions are very low and not likely to
cause any interference in nearby equipment.
RF emissions
CISPR 11
Class A
The ArthrexID Video System is suitable for use
in all establishments other than domestic, and may be
used in domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning:
This equipment/system is intended for
use by healthcare professionals only. This
equipment/ system may cause radio interference or
may disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures,
such as re-orienting or relocating the
ArthrexID Video System or shielding the location.
.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
System Cables
Type
Use
Shielded?
Ferrite?
Maximum Length
Power Cord
Supply Line Power to
the Box
No
No
2 M
BNC to BNC
Composite Video Out to
Monitor
Yes
No
1.8 M
Display Port
Display Port Output to
Video Monitor
Yes
Yes
2.0 M
DVI
DVI Video Out to
Monitor
Yes
Yes
2.0 M
3 pin mini Stereo
Accessory Port
Yes
No
1.8 M
Ethernet
CCU to Computer
Yes
No
5 M
USB
CCU to Printer
Yes
No
1.8 M