2.0
System Installation
and Operation with
Data Input Device
2.1
Installation
1. Your Synergy
ID
Camera Control Unit will
indicate which software configuration is
enabled at boot up on the Video Monitor’s
Splash screen.
2.1.1
Typical System Installation
NOTE: See Typical Interconnect Diagram,
NOTE: Synergy
ID
may be installed in a tower
or on an equipment boom.
1. Place Synergy
ID
Camera Control Unit (CCU)
on tower shelf or installed on equipment
boom.
2. Attach monitor to the tower or equipment
boom and connect monitor DC power cable
to the rear panel of the monitor as shown.
3. Attach Synergy
ID
Data Input Device to
secondary tower arm or equipment boom.
Connect the cable from the Data Input
Device to the connector labeled “tablet” on
the back of the Synergy
ID
CCU.
4. Connect a Display Port cable to a Display
Port output on the rear panel of the
Synergy
ID
CCU. Connect the other end of
the Display Port cable to the Display Port
input of the display monitor. (3G-SDI or DVI
cables may be used instead of Display Port
cables.) Note: Arthrex recommends
connecting Synergy
ID
to the primary surgical
monitor via multiple output types (e.g.,
Display Port and DVI, Display Port and 3G-
SDI) in the event that one type of connection
is lost.
5. If using a printer, connect the printer cable to
the USB connector on the rear panel of the
Synergy
ID
CCU. Connect the other end of
printer cable to the printer.
6. If using a Synergy Laser Light Source,
connect the laser light source cable from the
laser light source to the connector labeled
“Laser Input” on the back of the Synergy
ID
CCU.
7. Plug the AC power cord into the Synergy
ID
power inlet module and a standard
grounded AC Mains outlet (100-240 V˜, 50-
60Hz).
8. Insert the card edge connector of the
Synergy
ID
camera head into the camera
receptacle on the front of the CCU. Ensure
the camera had connector contacts are clean
and dry prior to insertion.
WARNING: Inserting an incompatible
Camera Head into the camera receptacle
(see Figure 1) can result in damage to the
CCU.
8. Connect the Light Guide cable into the Light
Guide receptacle on the front panel of the
Synergy
ID
CCU. Attach the other end of the
Light Guide cable to the endoscope. If using
a bifurcated light guide, connect the
additional bifurcated light guide connect to
the Synergy Laser Light Source.
9. Insert the endoscope into the Synergy
ID
camera head grasping mechanism or into
the C-Mount Coupler for C-Mount Heads.
10. Press the LED Light Engine On/Standby
Switch to activate the LED light engine.
NOTE
: If there is no Light Guide cable
connected to the Synergy
ID
CCU, pressing
the On/Standby Switch will not activate the
LED light engine until one is connected.