5.1.1 Instrument [Warning] Causes and Remedies
A warning occurs as a result of an incorrect operating procedure. You can
continue measurement by following simple remedial procedures.
No.
Description
Cause
Remedy
[W-001]
The program has
been changed.
The program has been
changed due to the upgrade
of ROM, etc.
Tap [OK].
The instrument will
automatically perform
necessary processing.
[W-002]
Remove the test
strip.
The power cannot be turned
off because the test strip is
left on the test strip holder.
Remove the test strip and
tap [OK].
[W-003]
The test strip
holder is not
attached properly.
The test strip holder is
attached improperly.
Properly attach the test
strip holder and tap [OK].
[W-011]
The battery level
is low.
Measurement cannot be
performed due to low
battery level.
Tap [OK].
Start measurement when
the battery has been
charged.